The Runyes RAY68(W)-2.0-800 AC Wall-Mounted Dental X-Ray Machine is a professional intraoral diagnostic radiography system designed for dental clinics requiring a fixed, space-efficient X-ray solution.
The RAY68(W) family uses Runyes' industrial-frequency AC X-ray platform with an original Japan-imported Canon (Toshiba) tube head. Published Runyes literature specifies a 70 kV tube voltage, 7 mA tube current and 0.7 mm focal spot.
The 2.0 generation adds Runyes' newer timer/control system with an integrated operating system. Microprocessor-controlled timing provides precise exposure control, while the digital display can provide fault-code indications and preserve parameters for digital imaging workflows. Runyes states that the system is compatible with digital dental sensors.
The -800 configuration differentiates this SKU from the standard RAY68(W) 2.0 and the longer RAY68(W) 2.0-1000 configuration. Current Runyes distributors separately list all three versions.
The articulated positioning system provides substantial movement around the dental chair. Runyes documentation states that the tube head can rotate approximately 370° horizontally, allowing positioning for different intraoral projections.
Key Features
AC intraoral dental X-ray system
RAY68(W) 2.0 series
800 mm configuration
Wall-mounted installation
Original Canon (Toshiba) tube head
70 kV ±10% tube voltage
7 mA ±20% tube current
0.7 mm focal spot
16° anode angle
0.06–2.0 second exposure range
900 VA maximum power
AC 230 V, 50/60 Hz
≥2.1 mm Al inherent filtration
Half-value layer of 1.6 mm Al at 70 kV
Microprocessor-controlled exposure timing
Digital display
Fault-code display function
New-generation integrated timer/control system
Articulated positioning arm
Tube-head horizontal rotation up to approximately 370°
Suitable for compatible digital intraoral sensors
Space-saving wall-mounted design
Technical Specifications
Specification Details
Brand Ningbo Runyes
Model RAY68(W)-2.0-800
Equipment Type Intraoral Dental X-Ray Machine
Technology AC / Industrial Frequency
Installation Wall-Mounted
Configuration 2.0 – 800 mm
Tube Head Canon (Toshiba)
Focal Spot 0.7 mm
Tube Voltage 70 kV ±10%
Tube Current 7 mA ±20%
Anode Angle 16°
Exposure Time 0.06–2.0 seconds
Power Voltage AC 230 V
Frequency 50/60 Hz
Maximum Power 900 VA
Electric Current 4 A
Fuse 6.3 A
Inherent Filtration ≥2.1 mm Al
Half-Value Layer 1.6 mm Al at 70 kV
Leakage Radiation ≤0.14 mGy/h at 1 m
Control Microprocessor / Integrated Digital Timer
Tube Head Rotation Approx. 370° horizontal
Application Intraoral Dental Radiography