NEWSKU-36-MTVO3OLE-AGID
Agappe Mispa i3 Automated Specific Protein Analyzer
Unit Price · Ex-Lagos
₦5,100,000
approx. $3,333.33 at today's rate
10 units available · MOQ 1 unit · Ships in 48hrs
× ₦5,100,000
Buyer protected Nationwide logistics Paystack verified
Fulfilled by Store
HH-00016
The Agappe Mispa i3 is a compact automated cartridge-based specific protein analyzer developed to simplify immunoprotein testing in clinical laboratories, hospitals and diagnostic centres.
It uses Nephelometry and Turbidimetry measurement technologies and Agappe's cartridge-based workflow to provide automated sample processing. Its Inject Aspiration System (IAS) aspirates the required sample and reagent volumes, helping reduce manual pipetting while the system performs functions including dilution, washing, mixing, incubation and measurement.
The Mispa i3 features a 7-inch colour TFT touchscreen, smart-card calibration, 650 nm laser LED light source, built-in thermal printer, optional handheld barcode reader and storage for 1,000 patient results. The minimum sample volume is 80 µL.
Agappe offers a broad assay menu for the system, including tests such as CRP, ASO, RF, HbA1c, Microalbumin, Ferritin, Cystatin C, hsCRP, immunoglobulins, D-Dimer, Vitamin D and Procalcitonin (PCT), depending on the compatible reagent cartridge used.
Key Features
Automated cartridge-based specific protein analyzer
Nephelometry and Turbidimetry measurement
Prefilled, test-specific reagent cartridges
Inject Aspiration System (IAS)
Automatic sample dilution
Automatic mixing and incubation
Automatic washing
Smart Card calibration
650 nm laser LED light source
Controlled reaction temperature of 37°C ± 0.2°C
7-inch colour TFT touchscreen
Built-in thermal graphic printer
Optional handheld barcode reader
Stores up to 1,000 results
Minimum sample volume of 80 µL
RS-232 computer interface
Compact benchtop design
Wide specific-protein assay menu
Detailed Technical Specifications
Specification Agappe Mispa i3
Analyzer Type Automated Specific Protein Analyzer
System Cartridge-Based
Test Method Nephelometry & Turbidimetry
Light Source Laser LED, 650 nm
Laser Power 5 mW
Photodetector Silicon Diode
Temperature Control Dry Heating
Reaction Temperature 37°C ± 0.2°C
Calibration Smart Card Calibration
Display 7-inch Colour TFT
User Interface Touchscreen
Minimum Sample Volume 80 µL
Result Storage 1,000 Results
Printer Built-in Thermal Graphic Printer
Barcode Reader Handheld Reader – Optional
PC Interface RS-232
Power Supply 85–265 VAC, 50/60 Hz
Power Consumption 100 W
Dimensions (L × W × H) 380 × 353 × 352 mm
Weight 7.5 kg without battery
Reagent System Prefilled IVD Reagent Cartridges